Filmomat CLASSIC - Maintenance & Inspection Guide
Regular inspection and maintenance help ensure reliable operation and protect your film from damage. Please follow the recommendations below.
Before Every Processing Run / Start of the Day
1. Check the Tank Drive
Attach the film tank cap.
Using Manual Control, briefly start the tank rotation.
Verify that:
The drive disk rotates smoothly.
The drive disk is securely attached to the motor shaft and does not wobble or slip.
2. Inspect the Film Tank
Ensure the inside of the film tank is clean and free of any obstructions.
Inspect the tank connector screw. Although uncommon, it can loosen over time and protrude into the tank, preventing film spirals from rotating freely.
Important: A malfunctioning tank cap or blocked film spiral can result in severe film damage.
During Every Processing Run
Monitor Water Circulation
Confirm that the water pump is circulating the water bath continuously.
Insufficient water circulation can cause:
Incorrect processing temperatures
Overheating of the heating element
Monitor Chemical Pumping
Verify that all processing chemicals are pumped into and out of the film tank correctly.
Watch for abnormal flow or incomplete filling/draining during each processing step.
Daily Maintenance
After the final processing run of the day:
Run the Cleaning Cycle.
Drain all liquids from the machine.
Wipe the interior dry with a clean paper towel to remove any remaining moisture or chemical residue.
Preventive Maintenance & Troubleshooting
Loud Rattling Noise During Chemical Pumping
Symptoms
Loud rattling or grinding noise
Sudden loss of pumping power
Residual chemistry remains in the film tank or tubing
Likely Cause
Worn main pump impeller
Solution
Replace the main pump impeller. Contact manufacturer for guidance and spare parts.
As preventive maintenance, replace the impeller approximately every 500 processing runs.
FILM TANK IS HARD TO ATTACH / REMOVE
Symptoms
Attaching and removing of the film tank is very hard.
Likely Cause
Worn tank connector seals
Solution
Lubricate the seals with a tiny drop of oil or grease.
If the seals are worn down a lot, replace them with new ones (NBR, 20×2mm)
Water Bath Is Not Circulating
Symptoms
No visible water movement
Likely Cause
Defective or blocked water circulation pump
Solution
Contact the manufacturer for troubleshooting and repair instructions.
Cross Contamination Between Chemical Compartments
Symptoms
Activating one compartment (A, B, C, or Water) also pumps liquid from another compartment.
Likely Cause
Solenoid valve not closing properly due to debris or worn valve seals
Solution
Inspect the solenoid valve seals for contamination or wear. Contact the manufacturer for service instructions.
Machine Shuts Down When Heating Starts
Symptoms
The processor powers off immediately after the heating cycle begins.
Likely Cause
Defective heating element
Solution
Contact the manufacturer to obtain a replacement heating element and installation guidance.